Good Laboratory Practice
Sample of the GLP Test Report
GLP Test Report
BRUKER BIOSPIN GMBH, NMR-DEPARTMENT | |
Company/Institution: | Bruker BioSpin GmbH |
Name customer: | Marcel Burgers |
Street: | Silberstreifen |
Town: | Rheinstetten 76287 |
Country: | Germany |
Name operator: | Mike Engelhardt |
Phone (Lab): | +49 (0)721-5161-145 |
Fax: | +49 (0)721-5161-297 |
ISDN: |
|
E-mail: | mbu@bruker.de |
Spectrometer type: | AVANCE400 |
System order #: | HH001234 |
Contact Bruker BioSpin AG: | Mike Engelhardt |
Phone Bruker BioSpin AG: | +49 (0)721-504-440 |
Fax Bruker BioSpin AG: | +49 (0)721-504-480 |
Email Bruker BioSpin AG: | eng@bruker.de |
Email Bruker BioSpin AG: |
|
ISDN Bruker BioSpin AG: |
|
Sales contract #: |
|
Service contract #: |
|
| |
Result of loop over all experiments | |
Lineshape test for 1H with rotation | OK |
Resolution test for 13C | OK |
Water suppression test | OK |
| |
Experiment: Lineshape test for 1H with rotation | |
Sample: 1% CHCl3 | |
Specification result of humpcal: | OK |
Line width at 0.55% of signal height = 7.2 Hz ( < 10Hz) | OK |
Line width at 0.11% of signal height = 18.7 Hz ( < 20Hz) | OK |
Half width = 0.31 Hz ( < 0.4Hz) | OK |
Dataset name: GLP.980626 2 1 v demo | |
Experiment: Resolution test for 13C | |
Sample: ASTM | |
Specification result of hwcal: | OK |
Half width = 0.08 Hz ( < 0.1Hz) | OK |
Dataset name: GLP.980626 3 1 v demo | |
Experiment: Water suppression test | |
Sample: 0.002M SUCROSE | |
Specification result of suppcal: | OK |
Line width at 50% of DDS signal = 38.3 Hz ( < 40Hz) | OK |
Line width at 10% of DDS signal = 75.1 Hz ( < 80Hz) | OK |
Splitting = 32 % (< 36%) | OK |
Sensitivity = 86.20 :1 ( 64:1) | OK |
| |
Date: |
|
Test performed by: | |
Test results approved by: _______________________________________ | |
Signature: ___________________________________ Date: ___________ | |