Language
Search
Product News
- Bruker Introduces HyperQuant™, a Unique Bench-Top NMR Reader to Quantify Hyperpolarization
- Bruker BioSpin's MRI CryoProbe(TM) Selected for 2008 R&D 100 Award for Enabling Preclinical MRI on...
- Bruker BioSpin 1.7mm MicroCryoProbe(TM) Wins R&D 100 Award for Delivering a Tenfold Jump in NMR...
- NMR Newsletter 07/2008
- ParaVision 5.0 released- Routine Workflow for Preclinical MRI
BGA-S Gradient Series
High Performance Integrated Gradient and Shim Systems
An increasing number of preclinical MR imaging applications demand ultra-high magnetic fields with increased requirements to gradient and shimming technology. Bruker BioSpin MRI therefore is introducing the BGA-S series, a complete new series of high performance actively shielded gradient systems with integrated shim coils.
Superior performance on the following major characteristics can thus be achieved:
- Ultra high gradient strength
- Very short gradient slew rates
- Excellent gradient duty cycle (independent of shims and gradients)
- Optimal gradient linearity
- Very high gradient shielding quality
- Maximal shim strength
The new BGA-S series is available for BioSpec®, PharmaScan® and ClinScan® systems.
Bruker BGA-S Gradient- and Shim Systems
| BGA6-S | BGA9-S | BGA12-S | BGA20-S | |
| Outer diameter (mm) | 113 | 150 | 206 | 303 |
| Inner diameter (mm) | 60 | 90 | 116 | 200 |
| Strength* (mT/m) at Imax | 1000 | 740 | 660 | 300 |
| Slew rate* (T/m/s) at Umax | 11250 | 6600 | 4570 | 1100 |
| Gradient linearity/DSV (+/- %/mm) | 5/35 | 5/60 | 4/80 | 3/130 |
| Number of RT Shims | 9 | 9 | 9 | 9 |
| Max. cont. gradient all axis (mT/m) | 230 | 190 | 230 | 87 |
| Max. cont. gradient one axis (mT/m) | 350 | 210 | 180 | 60 |
| Umax (V)/Imax (A) | 300/200 | 300/200 | 500/300 | 500/300 |
* Output values have been measured on MRI systems